SC Series Instruction Manual - Model SC500 page 129
Error Me
ssage
s
15
Restart the instrument.
Error 30,
S.EE ERASE NAK
The channel’s EEPROM could not be detected during an erase operation.
Restart the instrument.
Error 31,
BLANK IDENT
The channel does not appear to have been installed properly. This occurs if you
swap Input or Output boards without using
SYSTEM MENU
->
INSTALL CHANNEL
to format
the channel’s EEPROM.
Error 32,
BOGUS EE ADDR
An invalid address has been passed to the programmer. Restart the instrument. If
this problem persists, contact our Service department with the information found
in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 33,
SIG.MOD WIRES
The instrument has detected that the (+)Signature and (-)Signature pins (labeled
as “+MEM” and “-MEM) are likely reversed.
Error 34,
SIG.MOD MISMATCH
The excitation type byte found inside a
Signature Module
indicates that the chan-
nel is not designed to operate with that type of transducer.
Error 35,
BOGUS MENU TYPE
A menu contains an invalid menu type. Contact Sensotec with the information
found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 36,
BAD EE TABLE
The instrument doesn’t understand the EEPROM type setting for the channel.
Contact Sensotec with the information found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 37,
BAD PERSONALITY
The channel doesn’t understand the given card type. Contact Sensotec with the
information found in the
SYSTEM MENU
->
SOFTWARE REV.
menu item.
Error 38,
BAD FORMAT
The channel’s EEPROM memory contains a display setting which is not under-
stood by the instrument. Check that every item in the channel’s
DISPLAY SETUP
menu has a valid setting.
Error 39,
WRONG HARDWARE
The channel type expected by the instrument for the channel number is not the
channel type that is physically installed. If you have changed the address jumpers
on one or more channels, replace them in their original configuration.
Error 41,
SHUNT CAL ERROR
The application of the shunt resistor did not affect the transducer’s output. Check
the wiring and the transducer output with a voltmeter to verify proper operation.
If this error occurs on a Strain-Gage Channel:
•
Verify that the shunt calibration resistor installed on the Main Circuit Board is
the correct value per the transducer’s Calibration Record.
•
Most unamplified strain-gage transducers do not include a “buffered shunt
cal” feature. Verify that jumper settings are correct according to Table 4-8,
“Shunt Cal Application Configuration,” on page 54.
If this error occurs on a High-Level Input Channel, make certain that the Shunt 1
and Shunt 2 pins are connected to the correct pins of the transducer/amplifier.
Error 43,
BAD IDENT CRC
A checksum failure has occurred when reading a channel’s EEPROM. Restart
the instrument.
Error 44,
BAD FUNCTION
The instrument has requested that the channel perform a function that it is not
capable of. Restart the instrument.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...