10.3 Calibration Procedure
If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page 23. A listing of all menu items is given in
“Setup Menu Reference” on page 133.
Step 1: Wire the transducer to the channel’s connector.
See “Wiring” on page 52 for details.
Step 2: Enter the
CALIBRATION TYPE
.
There are three methods that can be used to calibrate the transducer to the
Input Channel. Each has advantages and disadvantages as described in
“
Menu Item” on page 60. It is important to know your
application in order to select the appropriate calibration type.
Step 3: Enter the
CALIBRATION DATA.
If your transducer has
Signature Calibration
, you don’t need to enter these
values as they are entered automatically.
If your transducer was ordered and shipped along with the instrument, you
don’t need to enter these values as they have been entered at the factory.
Otherwise, consult the Certificate of Calibration for the transducer when
entering information in the
CALIBRATION DATA
sub-menu.
Step 4: Perform the calibration.
Otherwise, use the
CALIBRATE
menu item to start the calibration process. You
will be prompted to apply loads to the transducer as required.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...