page 56 008-0671-00
10.4 Specifications
TRANSDUCER INPUT
Transducer type
full-bridge, strain gage
Excitation Supply
5 VDC @ 60 mA max.
Excitation Overcurrent Protection
yes
Transducer full-scale output
0.5 mV/V to 21.0 mV/V
Amplifier Gain Selection
automatic
Calibration Type
shunt, mV/V or
2-, 3- or 5- point known load
Differential Input Voltage
+/- 105 mV (max.)
A/D Converter
24-bit Sigma-Delta
Low-pass filter
digital, 24-tap FIR
Resolution and Frequency Response
INSTRUMENT-ONLY ACCURACY
(Frequency Response setting<=16Hz; Linearity, repeat-
ability & hysteresis)
mV/V Calibration
±0.1% F.S.
Known Load Calibration
±0.01% F.S.
AUXILLIARY INPUTS
Quantity
2
Type
momentary contact closure
Response Time
< 5ms
Field-Selectable Functions
tare on, tare off,
peak/valley clear, peak/valley hold,
track hold
DIGITAL-TO-ANALOG OUTPUT
Availibility
with Option 58i
Output voltage range
5, ±5, 10 or ±10 VDC (field selectable)
Source
any channel’s track, peak or valley value
Isolation
500V
Resolution
14 bits
Frequency Response
same as input when driven by the same
channel’s tracking data
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...