SC Series Instruction Manual - Model SC500 page 67
Str
ain
-G
ag
e In
pu
t C
han
ne
l
10
10.5.8
SIGNATURE MODULE
Sub-Menu
UPDATE SIG.MOD?
Menu Item
Updates the information stored in the transducer’s
Signature Module
with any new
settings that is entered into the instrument with the SETUP menus. The settings
that will be updated are:
•
EXCITATION
•
FULL SCALE MV/V
•
FULL SCALE VALUE
•
SHUNT CAL VALUE
•
DISPLAY. UNITS
Note that changes made to the instruments display setup (such as decimal point
and count-by settings) are not updated into the
Signature Module
.
Updating enters these new settings from the instrument into the
Signature Mod-
ule
. If you do not update the
Signature Module
, connect a different transducer to
the instrument and then re-connect the original transducer again, the instrument
will revert to the settings that are stored inside the
Signature Module
.
The options for this menu item are:
•
“
NO
” will not update the
Signature Module
.
•
“
YES
” will display
WORKING
, update the
Signature Module
, then display
DONE
.
DEFAULT SIG.MOD?
Menu Item
This menu item will overwrite any user settings in the
Signature Module
with the
original default settings. The
Signature Module
always maintains a copy of the
original default settings that were created at the factory.
The options for this menu item are:
•
“
NO
” will not restore original default settings.
•
“
YES
” will display
WORKING
, update the
Signature Module
, with the original factory
settings, then display
DONE
.
UPDATE DAC SETUP
Menu Item
This menu item determines whether or not the
DAC. FULL SCALE
and
DAC. ZERO SCALE
settings are overwritten with information from a Signature Module when a new
transducer is detected by the instrument.
The options for this menu item are:
•
“
YES
” will allow a newly detected
Signature Module
to update the DAC Output
settings.
•
“
NO
” will not allow automatic updating of the DAC Output settings.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...