12.2.8 Low Voltage DC-DC
LVDTs
Use the following wiring diagram when connecting a low-voltage DC-DC LVDT to
a High-Level Input channel. Examples of such devices include
•
Models DLB, DLE and DLF DC-DC LVDTs
The High-Level Input channel’s Configuration Jumpers must be set as follows for
proper operation. See “Excitation and Signal Jumpers” on page 100.
•
(+)Excitation supply: “+12 VDC”
•
(-)Excitation supply: “GND”
•
Signal type: “voltage”
•
Signal reference: “single ended”
Figure 12-7: “Low Voltage” DC-DC LVDT Connection to High-Level Input Channel