2.4 Use the SETUP Menus to Enter Transducer Information
You can skip this step if:
•
You are using a transducer equipped with
Signature Calibration
, or
•
The transducer and instrument were purchased together and set up by Sen-
sotec.
Otherwise, you must enter information about your transducer into the SETUP
menu of the channel to which it is connected. See the appropriate chapter of this
manual for that card type.
2.5 Calibrate the Transducer to the Input Channel
You can skip this step if:
•
You are using a transducer equipped with
Signature Calibration
, or
•
The transducer and instrument were purchased together and set up by Sen-
sotec.
Otherwise, you must choose a calibration method (e.g. Shunt Calibration, mV/V
Calibration, or Known Load Calibration) appropriate for your application and use
the SETUP menu for the Input Channel (“
CHANNEL 01 MENU
”) to calibrate the channel
to the transducer. See the appropriate chapter of this manual for the input chan-
nel type that was purchased with the instrument.
2.6 The SC Series Instrument is Ready for Use
See “RUN Mode” on page 20 for information on how to operate the instrument
while it is in the RUN mode.
Apply some test stimulus on the transducer to observe changes in the display.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...