page 34 008-0671-00
5.4 Relay Output Pins
5.4.1 Operation
When one of the Limits is activated, the corresponding Relay NO pin on the Options
connector will be connected to the corresponding Relay COM pin.
5.4.2 Specifications
5.4.3 Diagnostics
Sub-Menu” on page 35 for information on how to manually activate
and deactivate each of relay pins from the SETUP Menu.
NOTICE
A SensoCode program running on a Mathematics Virtual Channel may over-
ride the default behavior of the relay outputs. Consult the Customer Information
Sheet included with your instrument for details.
RELAY OUTPUT
Quantity and Type
2 form C (Option 58a) or
3 form A (Option 58h)
Contact Rating
0.5A @ 50VAC max.
1A @ 30VDC max.