SC Series Instruction Manual - Model SC500 page 83
AC
/A
C-L
VDT Input Channel
11
11.5.6
CALIBRATE
Menu
Item
This menu item performs a calibration according to what was entered in the
CALI-
BRATION TYPE
and
CALIBRATION DATA
menu items.
If the
CALIBRATION TYPE
is
2-Point Known Displacement Calibration
...
•
The display will read
DOING 2POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where 00000 and UNIT are previously entered as
KNOWN POINT 1/2
and
DISPLAY.
UNITS
). When you have applied this displacement to the transducer, press
[ENTER].
•
The display will read
WORKING
, then
APPLY 01000 UNIT
(where 01000 and UNIT are
previously entered as
KNOWN POINT 2/2
and
DISPLAY. UNITS
). When you have
applied this displacement to the transducer, press [ENTER].
•
Upon completion, the display will indicate
DONE
and the instrument will return to
the RUN mode.
If the
CALIBRATION TYPE
is
3-Point Known Displacement Calibration
...
•
The display will read
DOING 3POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where 00000 and UNIT are previously entered as
KNOWN POINT 1/3
and
DISPLAY.
UNITS
). When you have applied this displacement to the transducer, press
[ENTER].
•
The display will read
WORKING
, then
APPLY 00500 UNIT
(where 00500 and UNIT are
previously entered as
KNOWN POINT 2/3
and
DISPLAY. UNITS
). When you have
applied this displacement to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 01000 UNIT
(where 01000 and UNIT are
previously entered as
KNOWN POINT 2/3
and
DISPLAY. UNITS
). When you have
applied this displacement to the transducer, press [ENTER].
•
Upon completion, the display will indicate
DONE
and the instrument will return to
the RUN mode.
If the
CALIBRATION TYPE
is
5-Point Known Displacement Calibration
...
•
The display will read
DOING 5POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where 00000 and UNIT are previously entered as
KNOWN POINT 1/5
and
DISPLAY.
UNITS
). When you have applied this displacement to the transducer, press
[ENTER].
•
The display will read
WORKING
, then
APPLY 00250 UNIT
(where 00250 and UNIT are
previously entered as
KNOWN POINT 2/5
and
DISPLAY. UNITS
). When you have
applied this displacement to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 00500 UNIT
(where 00500 and UNIT are
NOTICE
Before performing a calibration,
the transducer must be connected to the instrument, the
CALIBRATION TYPE
must
be selected (see “
Menu Item” on page 80) and the
CALIBRA-
TION DATA
must be entered (see “
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...