12 General specifications
Signalpoint Pro Detector
Use
Fixed point gas detector designed to detect toxic or Oxygen gas hazards that are
commonly found in industrial applications. When installed with a suitable barrier it is
suitable for safe area and European Zone 1 or 2 and North American Class I Division 1
areas.
Detectable Gases
1
Electrical
Connections
2 wire loop powered (source)
Input Voltage Range:
14 volts min. to 28 volts max., DC
Max Power
Consumption:
Less than 1.0 W
Output Current range:
3mA. to 23mA
Max loop resistance:
Refer to section 6.3
Loop Distance (MAX):
1km (3000ft) using 1mm
2
CSA cable or equivalent.
Barrier requirements:
Ca > Ci+C, La > Li+L, Voc<Vmax=28V, Isc<Imax=125mA.
Recommended Cable
2 wire with screen
0.5mm
2
(20AWG) to 1mm
2
(16AWG)
Signal
0-100% FSD 4-20mA
Max. over range 23mA
4mA (for toxic) and 17.38mA (for O
2
) auto inhibit during calibration
Fault
≤
3mA
Construction
Material
Grey ABS/PPS
Maximum Dimensions
150 x 105 x 83.5mm (5.9 x 4.1 x 3.3”)
Weight
479g (15.4oz.)
Entries
1 x M20 clearance (LHS), 1 x M20 clearance knockout (RHS)
Environmental
IP Rating
IP66 as standard suitable for use in and out of doors (EN 60529:1991 / A1:2001)
Operating Temperature
-20°C to +55°C (-4°F to 131°F)
Operating Humidity
Continuous 20-90% RH (non condensing)
Operating Pressure
90-110kPa
Storage Conditions
15°C to 30°C (59°F to 86°F) 30-70% RH (non condensing)
Approvals
CE compliant in accordance with:
EMC Directive 89 / 336 / EEC as amended by 92 / 31 / EEC
EN50270 Type 2 Heavy Industrial for susceptibility
EN55011B Light Industrial for emissions
Certification
US and Canadian:
European:
Intrinsically Safe; Class I, Division 1, Groups A, B, C, D, E, F & G
Intrinsically Safe:
II 2 G Ex ia IIC T4
1
Detection performance is temperature and humidity dependent. Listed data is based at 20°C, 50% RH only. Response time is
longer when operating in colder temperatures.
* Based on a 20-25°C temperature, 50 to 80% RH, using a calibration gas flow housing on a freshly calibrated instrument. If different
cylinders are used, other that the calibrating source, the cylinder tolerance will be taken into consideration.
(*) -20°C to +40°C; 20% from +40°C to +55°C
(**) 0 to +40°C, 30% from -20°C to 0°C and +40°C to +55°C
(***) -10°C to +40°C, 30% outside of -10°C to +40°C
(****) Over temp. range -20°C to +40°C
(
#
) +10.0 to +50.0 ppm
Less than +10.0ppm
+/-10% -10°C to +40°C
+/-20% less than -10°C
+/-35% more than +40°C
Do not use the weather proof cap with a SO
2
or NO
2
sensor.
Note: the calibration gas level used must be between 30% and 70% of the full scale level for best accuracy.
Oxygen
0.0-25.0% Vol
25%VOL only
20.9%/Vol (Fixed) 20.9%/Vol -20 to 55 deg. C
15
3% FSD
n/a
Hydrogen Sulfide
0-15.0ppm
10.0 to 50.0ppm (1.0) 5.0 to 30.0ppm
10ppm
-20 to 55 deg. C
30
10% (*)
<1% FSD (****)
Hydrogen Sulfide
0-100ppm
50 to 500ppm (10)
20 to 300ppm
50ppm
-20 to 55 deg. C
30
10% (*)
<1% FSD (****)
Carbon Monoxide
0-300ppm
100 to 999ppm (100)
40 to 600ppm
100ppm
-20 to 55 deg. C
45
10% (*)
<3% FSD (****)
Sulfur Dioxide
0.0-15.0ppm
5.0 to 20.0ppm (5.0)
2.0 to 12.0ppm
7.5ppm
-20 to 55 deg. C
90
20% (**)
n/a
Ammonia
0-200ppm
50 to 200ppm (50)
20 to 120 ppm
100ppm
-20 to 55 deg. C
180
20% (***)
0
Ammonia
0-1,000ppm 200 to 1,000ppm (50)
80 to 600ppm
300ppm
-20 to 55 deg. C
210
20% (***)
0
Nitrogen Dioxide
0.0-10.0ppm
5.0 to 50.0ppm (5.0)
2.0 to 30.0ppm
5.0ppm
-20 to 40 deg. C
60
15% (#)
0
Hydrogen
0-1,000ppm
1,000ppm only
400 to 600 ppm
500ppm
-10 to 40 deg. C
90
25%
0
Gas
User Selectable
Gas Range
(Step Value)
Selectable Cal
Gas Range
Default Gas
Range
Zero Drift
(+/-)
Default
Cal Point
Operating
Temp. Range,
deg. C
Response
Time* (T
90
)
secs
Ac/-
(%FSD or % of
applied gas)