IMPORTANT!
LED Sensors are light sensitive. Use Hopkins® Y-Probe wrap on
top of the Adjustable Y-Probe Sensor with belt to better secure the probe.
Inspect the sensor site every 2 hours for skin integrity and relocate the sensor
every 2 hours according to steps 1-4 .
9 Specifications
Environmental
specifications
Operating conditions
Storage conditions
Temperature
5 to 40ºC(41 to 104ºF)
-20 to 60ºC(-4 to 140ºF)
Relative humidity
(non-condensing)
15% to 95%
15% to 95%
Hardware specifications
Display
Loudspeaker
Power indicating lamp
Alarm indicating lamp
Sensor connector
Power supply connecto
Color LCD, 2.4”
1.Gives audible alarm (45 to 85dB) and button tone;
Supports Pitch Tone and multi-level volume.
1.lighting green and yellow
1. lighting red and yellow
1.9-pin type D connector
1. used to connecting the Charger stand
Communication port
Data storage
Shelf life
Alkaline batteries
Quantity
Specification
Specification
Run time
Shutdown delay
No
No
5 Years
3
1.5 V, AA
2000 mAh
22 hours with SpO2 monitored continuously,audio
indicators off and backlight brightness set to minimum
using a new,fully charged battery at ambient
temperature 25
C
Max. 10 minutes after the low battery alarm first occurs
Measurement specifications
Spo2
Display Range
Accuracy
Resolution
1 to 100%
1%
70 to 100%: ±2% (measured without motion)
0% to 69%: Unspecified
-26-