DeltaMyc Operation Manual J810018 (7 Oct 2014)
Technical Specifications
6-2
Manual control
Through joystick
Automatic control
Through DataStation software
TCSPC electronics
System type
Detection via single-photon counting
Resolution
55 ps/channel with DeltaHub
Lifetime range
25 ps to 10 µs, depending upon sample
Detector
System type
PPD fast photomultiplier tube
Spectral range
250 to 650 nm or 300 to 850 nm
Transit spread time
200 ps
Dark count
< 80 cps
Filters
Available positions
Eight, manually selected, interchangeable cubes
Excitation filters
Optional 10 nm bandpass filters or short-pass filters
Dichroic beam filters
Optional wavelength or 50/50
Emission filters
Optional 40 nm bandpass filters or long-pass filters
Neutral-density filters
Manually inserted
Software
Data-acquisition
DataStation
TCSPC mapping
High-speed scan
Data-analysis
DAS6 software, including reconvolution
Operating system
Windows
®
XP or Vista
Dimensions
Length × width × height
140 cm × 90 cm × 80 cm
Electrical
CE Compliance
EMC: 2004/108/EC