background image

Measurement 

1.  Open the sensor cover.  

2.  Rinse the  sensor with DI – or tap 

water. 

3.  Dry the sensor with a tissue 

.  

.

 

 

Rinse again with some of the next 
standard solution for increased 
accuracy 

4.  Cover the complete sensor with 

sample solution. 
 

5.  Close the sensor cover. 

 

6.  Switch on the meter. 
 
The meter will start measurement immediately.   will 
appear once the value is stable. 

A.S. mode:

 If the value changes   disappears until 

new stability is reached (continuous measurement) 

A.H. mode:

 Holds the last value in the display. Press 

MEAS

 to do another measurement. 

 

Using Sampling Sheet B 
Using this sheet, the sensor can 
be covered with only 50 µL to 
100 µL sample. 

1.  Take the sheet with tweezers 
2.  Wipe across the sample 
3.  Follow the steps for normal measurement. 

 

Using Sampling Sheet B and holder 
Tiny particles in samples (e.g. extract from soil) 
influence the measurement result. Use the sampling 
sheet holder cover and sampling sheet B to 
counteract the influence. 

1.  Replace the sensor cover with the sheet holder 
2.  Open the sheet holder 
3.  Place a sampling sheet on the sensor 
4.  Close the sheet holder 
5.  Put a few drops of sample in the sheet holder 
6.  Follow the steps for normal measurement 

 

 

View mV or Temperature 
Only available in A.S. Stability mode, see Setting 
Menu. From measurement: 

1.  Press 

MEAS

 to view temperature 

2.  Press 

MEAS

 again to view mV 

3.  Press 

MEAS

 again to view units 

 

 

 

Reset to factory defaults 

All settings will be reset to the factory default settings. 
Calibration data will be deleted. 

1.  Power OFF the meter. 
2.  Simultaneously press and hold 

MEAS

,

 ON/OFF 

and 

CAL

 keys (approx. 3 seconds). 

3.  Press 

CAL

 to confirm or 

ON/OFF

 to cancel 

 

Maintenance and Storage 

 

Clean the sensor after use with DI - or tap water. 

 

Gently remove moisture from the sensor by 
dapping with a tissue. 

 

Close the sensor cap and slide cap when storing. 

 

Sensor can be stored dry. Perform sensor 
conditioning when the meter has be out of use for 
a longer period. 

 

Store at room temperature. 

 

 

Specifications 

 

 
 
 

 

Replacement sensors 

Part 

Art. number 

Description 

S022 

3200459867 

Sodium (Na) sensor for 
B-722 and NA-11 

S030 

3200459868 

Potassium (K) sensor 
for B-731 and K-11 

S040 

3200459870 

Nitrate (NO3) sensor for 
B-74x and NO3-11 

S050 

3200459869 

Calcium (Ca) sensor for 
B-751 and CA-11 

Standard solutions 6 x 14ml 

Y022H * 

3200457723 

Na Sodium 2000ppm 

Y022L * 

3200457724 

Na Sodium 150ppm 

Y041 

3200053433 

NO3 Nitrate 5000 ppm 

Y042 

3200053514 

NO3 Nitrate 300 ppm 

Y043 * 

3200053532 

NO3 Nitrate 2000 ppm 

Y044 

3200053535 

NO3 Nitrate 30 ppm 

Y045 * 

3200053536 

NO3 Nitrate 150 ppm 

Y031H * 

3200457719 

K Potassium 2000ppm 

Y031L * 

3200457720 

K Potassium 150ppm 

Y051H * 

3200457727 

Ca Calcium 2000ppm 

Y051L * 

3200457728 

Ca Calcium 150ppm 

Accessories 

Y-011A 

3014053435 

Sampling sheet C - 5 
Rolls, 11mm x 6m 

Y046 

3200053858 

Sampling sheet B - 
100pcs 

Y048 

3200459736 

Sampling sheet holder 
for LAQUAtwin series 

Crop 
Press 

3200254910 

Crop sap press 

 
* Solutions are included with the meter. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Features for all meters 

Measurement Principle 

Ion Selective Electrode 

Minimum Volume 

0.3ml (0.05ml with sampling sheet B) 

Resolution 

0 to 99 ppm: 1 ppm 

 

100 to 990 ppm: 10 ppm 

 

1000 to 9900 ppm: 100 ppm 

Calibration Points 

Up to 2 

Temperature Calibration 

5 to 40.0 ºC 

Temperature Display 

0 to 50.0 ºC 

Operating Temperature 

5 to 40.0 ºC 

Operating Humidity 

85% or less relative humidity (no condensation) 

Power 

CR2032 batteries (2) 

Battery Life 

Approx. 400 hours in continuous use 

Material 

ABS epoxy 

Dimensions 

164 x 29 x 20 mm (excluding projections) 

Mass 

Approx. 55 g (including sensor and batteries) 

Measurement modes 

Auto Hold / Auto Stable Measurement 

Automatic Power Off 

   After 8 minutes 

Low Battery Indicator 

   • 

IP67 Dust/Waterproof 

   • 

Replaceable Sensor 

   • 

Display 

LCD with backlight 

Warranty 

Meter 24 months / sensor 6 months 

Parameter 

CALCIUM  POTASSIUM  SODIUM 

NITRATE 

Model 

CA-11 

K-11 

NA-11 

NO3-11 

PART # 

3200689161  3200689160  3200689159 

3200689162 

Range ppm / 

mg/L 

4 to 9900 

2 to 9900  6 to 9900 NO3 / 1.4 to 

2200 NO3-N 

Range mmol/L 

0.1 to 250 

0.1 to 430   

Range kg/10a  

 

2 to 5000 

 

0.7 to 1100 

Accuracy 

±20% of actual 

value 

±10% of actual value 

Interfering ions 

and selective 

coefficients 

Fe

2+

, Zn

2+

: 1 

Fe

3+

: 10 

Cu

2+

: 1  10

-2 

(at 10

-3 

mol/L 

Ca

2+

)

 

Rb

+

: 1  10

-1 

(at 10

-3 

mol/L 

K

+

)

 

K

+

, Rb

+

110

-2 

(at 10

-

mol/L Na

+

)

 

I

-

: 10 

NO

2-

: 7  10

-1 

Cl

-

: 4  10

-2 

Br

-

: 9  10

-1 

(at 10

-3 

mol/L NO

3-

)

 

pH range 

4 to 12 pH (at 

10

-3 

mol/L Ca

2+

)

 

2 to 9 pH (at 

10

-3 

mol/L K

+

)

 

3 to 9pH (at 

10

-3 

mol/L 

Na

+

)

 

3 to 8 pH (at 10

-3 

mol/L NO

3-

)

 

Error  Solution 

Er1 

The meter may be defective.  

  Reset to factory defaults 

If resetting does not work the meter needs 
to be replaced. Repair is not possible. 

Er2 

Internal board is defective. 

  Meter needs to be replaced, repair is not 

possible. 

Er3 

Internal board is defective. 
Meter needs to be replaced, repair is not 
possible.

 

Er4 

Calibration error 

 

CAL

 is pressed in mV mode. 

  Temperature offset is too large, wait 

longer to perform temperature 
adjustment. 

Er4 + 

 

blinks 

Calibration error 

  Retry calibration 

  Ensure the standard solution complies 

with the set values for LOW and HIGH 
calibration points 

  Clean and recondition the sensor 

  Try new, fresh solutions 

  Clear Calibration data 

  Reset to factory defaults 

If the above does not work, the sensor 
might need replacement. 

Or or 
Ur 

Over or Under Range: value is higher (Or) 
or lower (Ur) than the meter’s limits 

  Measure a standard, if Or / Ur shows 

again, the sensor needs to be replaced. 

Tip 

HORIBA UK Limited 

Kyoto Close, Moulton Park, 
Northampton NN3 6FL 
Phone : 44 (0) 1604 542600 
Fax     : 44 (0) 1604 542696 
Email  : 

waterquality@horiba.com

 

www.horiba-laqua.com

 

 

 

Reviews: