Measurement
36
Setting temperature
error for FAIL
Specify the temperature error (E-21 to E-25) is output from the
FAIL contact or not.
yES, no
Setting no-flow error
for FAIL
Specify the no-flow error (E-61) is output from the FAIL contact
or not.
yES, no
FAiL output delay
time setting
Specify the FAIL contact output delay time in seconds after FAIL
status occurs.
Note:
The FAiL contact is immediately canceled without a delay.
0 to 600 (s)
Transmission output 1 setting
Transmission output 1
value (zero range)
setting
Specify the residual chlorine concentration value at the zero
point (4 mA) of transmission output 1.
0.00 to 5.00
(mg/L)
Transmission output 1
value (span range)
setting
Specify the residual chlorine concentration value at the span
point (20 mA) of transmission output 1.
0.00 to 5.00
(mg/L)
Transmission output
hold (current hold)
setting for
transmission output 1
Specify the transmission output value when transmission output
1 is in the hold mode.
HoLd: Holds and outputs the latest value
PrES: Outputs the preset value
HoLd,
PrES
Transmission output 1
hold output value
(preset) setting
Holds the preset value.
Note:
This is displayed when “Transmission output hold (current hold)
setting for transmission output 1” is set to PrES.
0.00 to 5.00
(mg/L)
Transmission output 1
burnout hold setting
Specify the output during burnout of the transmission output 1
when an error occurs during measurement.
out.4: Outputs 3.8 mA at the time of burnout
out.20: Outputs 21 mA at the time of burnout
non: No burnout setting
Note:
Transmission burnout appears only in the measurement mode.
non, out.4,
out.20
Transmission output 2 setting
Transmission output 2
value (zero range)
setting
Specify the temperature at the zero point (4 mA) of transmission
output 2.
10.0 to
110.0
(
C)
Transmission output 2
value (span range)
setting
Specify the temperature at the span point (20 mA) of
transmission output 2.
10.0 to
110.0
C)
Display
Item
Description
Setting
range