5 FUNCTION
92
5.9
Measurement Settings
The following are available for configuring measurement settings.
Displaying the setting screens for each measurement item
1.
Press the button (the following example: [turbidity]) of any of the measurement
items on the MEAS.SETTING screen.
The setting screen for the selected measurement item is displayed.
* This example is the TURB. screen.
Measurement
item
Setting
Description
Initial Setting
All
Measurement
Items
HIGH ALARM
(
LOW ALARM
(
UNIT HIGH ALARM
(
UNIT LOW ALARM
(
COEFFICIENTS
(
Even after calibration, measured values can deviate from their
actual values for a variety of reasons, so the coefficients can
help correct these values.
a=0
b=1.000
Turbidity
RANGE
(
Sets the measurement range.
0-4
Color
RANGE
(
Sets the measurement range.
0-20
TURB.
COEFFICIENTS
(
You can correct color when a measurement value deviates from
the actual value by using the turbidity.
0.000
Residual
chlorine
APPLIED VOLTAGE
(
Specifies the applied voltage.
50 mV
[MAINTE.]
MEAS. screen
MAINTENANCE screen
Displaying the maintenance item screen" (page 73).
[MEAS. SETTING]
MEAS. SETTING screen
Measurement item
(Seven items)
[TURB.], ...
[
]