5 FUNCTION
93
5.9.1
Setting the COEFFICIENTS
Reference
See "9.8 Evaluating Coefficients" (page 161).
1.
Press the COEFFICIENTS y=a+bx button on the TURB. MEAS. SETTING 1/2 screen
for each measurement item.
The TURB. COEFFICIENTS screen for all measurement items is displayed.
* This example is the TURB. screen.
2.
Press [a=x.xx].
The COEFFICIENTS (a) SETTING screen for the appropriate measurement item is
displayed.
3.
Enter the "a" value using the numerical key pad and press [SET].
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the right-most number of the value.
[MAINTE.]
MEAS. screen
MAINTENANCE screen
Displaying the setting screens for each measurement item" (page 92).
[(Measurement item)]
[MEAS. SETTING]
MEAS.SETTING
screen
[(Measurement item)]
MEAS.SETTING screen
COEFFICIENT
[a=x.xx b=x.xxx]
Measurement item
Input range
Turbidity
4.00 to 4.00
Color
20.0 to 20.0
Residual chlorine
2.00 to 2.00
Water pressure
1.000 to 1.000
pH (optional)
5.00 to 5.00
Conductivity (Optional)
50.0 to 50.0
Water temperature (optional)
25.0 to 25.0