5.12.4 Calibration history check
1.
Press [CALIBRATION HISTORY] on the DATA 1/2 screen.
The CALIBRATION HISTORY screen is displayed.
Maximum of 20 records of the calibration date or ZERO/SPAN CAL. are displayed in
chronologic order by measurement item or zero/span.
Reference
Calibration coefficient" (page 20)
* This example uses the TURB. screen.
[(Measurement items)]: Change the measurement items to display.
[SPAN]/[ZERO]: Change the display of zero calibration history and span calibration
history.
[
] [
]: Displays the next/previous page.
[Delete]: Deletes the calibration history.
Reference
To change the measurement item to display, see "
Changing the measurement items to display"
5.12.5 Deleting the calibration history
1.
Press [DELETE] on the CAL. HISTORY screen.
A confirmation screen to delete the calibration history is displayed.
2.
Press [YES].
The screen returns to the CAL. HISTORY screen.
Note
Leaves only the most recent zero/span calibration data, and deletes all of the calibration history
before it.
Displaying Data for an Item" (page 102).
[DATA]
MEASUREMENT screen
DATA screen
[CAL. HISTORY]
See"5.12.4 Calibration history check" (page 107).
MEASUREMENT screen
DATA screen
[CAL. HISTORY]
[DATA]
CAL. HISTORY screen