8 TROUBLESHOOTING
152
71
Water
temperat
ure
Concentration
lower limit
Sample water abnormality
Check sample water temperature.
Connection failure in water
temperature sensor
connector
Inspect water temperature sensor
connection.*
Water temperature sensor
failure
Replace water temperature
sensor.*
Internal analog circuit board
failure
Replace internal analog circuit
board (TW-ADC-01).*
72
Concentration
device upper
limit
Connection failure in water
temperature sensor
connector
Inspect water temperature sensor
connection.*
Water temperature sensor
failure
Replace water temperature
sensor.*
Internal analog circuit board
failure
Replace internal analog circuit
board (TW-ADC-01).*
73
Concentration
device lower
limit
Connection failure in water
temperature sensor
connector
Inspect water temperature sensor
connection.*
Water temperature sensor
failure
Replace water temperature
sensor.*
Internal analog circuit board
failure
Replace internal analog circuit
board (TW-ADC-01).*
75
Zero calibration
Mismatched calibration
results
Reset calibration value and
reexecute calibration.
Mismatching calibration
temperature
Check calibration temperature and
reexecute calibration.
Connection failure in water
temperature sensor
connector
Inspect water temperature sensor
connection.*
Water temperature sensor
failure
Replace water temperature
sensor.*
Internal analog circuit board
failure
Replace internal analog circuit
board (TW-ADC-01).*
76
Span calibration
Mismatched calibration
results
Reset calibration value and
reexecute calibration.
Mismatching calibration
temperature
Check calibration temperature and
reexecute calibration.
Connection failure in water
temperature sensor
connector
Inspect water temperature sensor
connection.*
Water temperature sensor
failure
Replace water temperature
sensor.*
Internal analog circuit board
failure
Replace internal analog circuit
board (TW-ADC-01).*
Alarm
No.
Item
Alarm
description
Cause
Corrective action
(*: Manufacturer service required)
Reference
Summary of Contents for TW-100