4 CALIBRATION
32
Span calibration <color>
Removing the residual-chlorine sensor
1.
Stop the device (close the sample water feed valve, discharge water and turn OFF
the power).
Reference
Stopping operation" (page 128).
2.
Remove the residual-chlorine sensor.
Reference
Removing the residual-chlorine sensor" (page 133).
3.
Attach the rubber plug to the residual chlorine/conductivity cell.
4.
Close the inner door of the device and turn ON the power.
Calibration operation
1.
Place the calibration tank with span calibration solution beside the device.
2.
Connect the tank to the calibration inlet of the device.
3.
Confirm that "WARM UP!" is not displayed.
If displayed, wait until it disappears.
Calibration
solution tank
Connect
Sample water feed valve
must be closed.
“WARM UP!”