5 FUNCTION
74
5.2
Input/Output Settings
You can configure the following settings related to input/output.
The I/O SETTING screen
1.
Press [I/O] on the SETTING 1/2 screen.
The I/O SETTING screen is displayed.
Setting
Explanation
Initial Setting
EXT. I/O IN
(
Select the method for external input.
However, you must configure the handling of abnormal water sampling
separately.
OFF
ALG. OUT (MAINTE.)
(
Select what kind of value is output when the device is in maintenance
mode
*1
.
HOLD
*2
ALG. OUT (SEQ.)
(
Select what kind of value is output when performing ACTION operations.
HOLD
*2
ALG. OUT(ALARM)
(
Select what kind of value is output when an alarm occurs.
HOLD
*2
WARMING UP
(
Select what kind of value to output when turning ON the power and
warming up.
Preset 4.0 mA
ANALOG OUTPUT
LEVEL
(
Selects the range of the analog output value.
4 mA to 20 mA
ALARM TIM
(
5 minutes
DISPLAYING MINUS
(
You can select to output the value with subtraction sign or to display a
blinking "0" when a measurement value is less than 0. This setting is useful
when you need to know that a measurement value is negative, but do not
need to know its exact value.
* This setting does not apply to water temperature (optional).
OFF
*1
Maintenance mode: The mode that the device is in when you press [MAINTE.] or [CAL]
and the MAINTENANCE or CAL. screen is displayed.
*2
HOLD:
Maintain the last measured value.
Displaying the maintenance item screen" (page 73).
[MAINTE.]
[SETTING]
MEAS. screen
MAINTENANCE screen
SETTING 1/2 screen
[I/O]
[
]