6 INPUT/OUTPUT
118
6
INPUT/OUTPUT
This device has interfaces for parallel input/output, analog output and RS-232C input/output
on the top of the device.
Use the I/O interfaces when you need to output information like warnings and measurement
values, as well as when you need to perform remote operation.
6.1
Analog Output
When outputting to an external device, this device uses an analog current for output.
Measurement items and analog output range, 4 mA to 20 mA
Measurement items and analog output range, 0 mA to 16 mA
Measurement item
Range
4 mA output
20 mA output
Turbidity
0 FTU to 2 FTU
0.00 FTU
2.00 FTU
0 FTU to 4 FTU
0.00 FTU
4.00 FTU
Color
0 CU to 10 CU
0.0 CU
10.0 CU
0 CU to 20 CU
0.0 CU
20.0 CU
Residual chlorine
0 mg/L to 2 mg/L
0.00 mg/L
2.00 mg/L
Water pressure
0 MPa to 1 MPa
0.000 MPa
1.000 MPa
pH (optional)
pH 2 to pH 12
pH 2.00
pH 12.00
Conductivity
(optional)
0 mS/m to 50 mS/m 0.0 mS/m
50.0 mS/m
Water temperature
(optional)
0
C to 50
C
0.0
C
50.0
C
Measurement item
Range
0 mA output
16 mA output
Turbidity
0 FTU to 2 FTU
0.00 FTU
2.00 FTU
0 FTU to 4 FTU
0.00 FTU
4.00 FTU
Color
0 CU to 10 CU
0.0 CU
10.0CU
0 CU to 20 CU
0.0 CU
20.0 CU
Residual chlorine
0 mg/L to 2 mg/L
0.00 mg/L
2.00 mg/L
Water pressure
0 MPa to 1 MPa
0.000 MPa
1.000 MPa
pH (optional)
pH 2 to pH 12
pH 2.00
pH 12.00
Conductivity
(optional)
0 mS/m to 50 mS/m 0.0 mS/m
50.0 mS/m
Water temperature
(optional)
0
C to 50
C
0.0
C
50.0
C