4 CALIBRATION
45
4.6.2
Individual zero calibration and span calibration <turbidity>
Preparation
Calibration tank…2
Zero calibration solution…1 L (Pour the solution into one of the calibration tanks)
Span calibration solution…1 L (Pour the solution into the other calibration tank)
Used fluid tank (Install it so that used fluid goes into the tank)
Calibration procedure
Perform calibration in the order: individual zero calibration
→
span calibration.
Individual zero calibration
1.
Place the calibration tank with zero liquid beside the device.
2.
Connect the tank to the calibration inlet of the device.
3.
Press the ZERO VALUE button ([x.xx]) on the TURB. CAL. screen.
The zero calibration value input screen is displayed.
MEASUREMENT screen CALIBRATION screen TURB. CAL. screen
See "4.3 Displaying the Calibration Screen" (page 23).
[CAL.] [TURB.]
Calibration
solution tank
Connect
Sample water feed valve
must be closed.
[x.xx]