5 FUNCTION
86
5.5.1
Setting the SAMPLING CHECK
1.
Press the SAMPLING CHECK button on the SAMPLING SETTING screen.
The SAMPLING CHECK screen is displayed.
2.
Select the settings using the buttons, and press [SET].
Returns to the SAMPLING SETTING screen.
5.5.2
Setting the SAMPLING TIME
1.
Press the SAMPLING TIME button on the SAMPLING SETTING screen.
The SAMPLING TIME SETTING screen is displayed.
2.
Enter the value of the water sampling time using the numerical key pad, and press
[SET].
Input range: 5 to 99 (min)
Note
Based on the water sampling time, select a container that is large enough to hold sample water.
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the right most number of the value.
Returns to the SAMPLING SETTING screen.
Displaying the SAMPLING SETTING screen" (page 85).
MEAS. screen
MAINTENANCE screen
SETTING1/2 screen
[SETTING]
[MAINTE.]
[SAMPLING SETTING]
SAMPLING SETTING
SAMPLING CHECK
[OFF]
/[ALARM ON]
/[EXT. I/O IN]
Setting
Explanation
OFF
Sample water is not obtained.
ALARM ON
When an alarm for abnormal concentration occurs
EXT. I/O IN
When given a signal to obtain an abnormal-water sampling from the contact input or serial input
Displaying the SAMPLING SETTING screen" (page 85).
MEAS. screen
MAINTENANCE screen
SETTING1/2 screen
[SETTING]
[MAINTE.]
[SAMPLING SETTING]
SAMPLING SETTING
SAMPLING TIME
[xx min]