Hospira GemStar System Operating Manual
147
Chapter 12. Troubleshooting
Alert Messages
Alert Display Message
Is the
pump
infusing?
Alarm Cause/Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
ALMOST EMPTY
Y
A/V
Delivery will complete in
30 minutes or less.
For bolus only program,
delivery will complete
during the next bolus
delivery.
•
Press [
STOP
] to clear
the alarm and put the
pump in STOP Mode.
•
Press [
SILENCE
] to
quiet the alarm.
•
No additional alarms
sound
Note: This alert is
available only with
Pain Management ther-
apy.
CHECK PRINTER
Y
A/V
•
The printer has not
responded to
commands from the
pump.
•
The printer is not
properly connected to
the pump.
1
Press [
SILENCE
] to
quiet the alarm for 2
minutes.
2
Ensure the printer is
properly connected to
the pump by checking
the following:
• Printer is
• powered on and in
ONLINE mode.
• Cable is plugged in
to both the pump
and the printer.
• Cable part number
and printer settings
are correct.
3
If the problem
continues, select
STOP PRINT
from
the Print Menu to
cancel all print jobs.
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...