Hospira GemStar System Operating Manual
157
Chapter 14
Operation Test
This test is designed to ensure the GemStar Pump is operating properly. Hospira
recommends performing this test a minimum of once every 12 months. Refer to
facility requirements to determine additional testing needs.
The test can be performed when the pump is in the
STOP
mode and requires
approximately 15 minutes to complete. The pump provides step-by-step instructions
through each section of the test.
Equipment Required
6
GemStar Pump
6
Two GemStar
Microbore
Pump Sets
without
a filter
6
Reservoir with at least 50 mL of water
6
20+ mL or larger graduated cylinder (graduated in 0.1 mL)
6
Two fresh, disposable AA batteries
6
Appropriate GemStar AC mains adaptor or other Hospira approved external
power source
d
Note: This test is available only when the pump is
UNLOCKED
.
WARNING: Disconnect the pump from the patient BEFORE
performing the Operation Test.
Summary of Contents for GemStar Series
Page 9: ...Notes viii Hospira GemStar System Operating Manual ...
Page 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Page 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Page 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Page 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Page 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Page 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Page 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Page 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Page 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Page 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Page 221: ...Notes 212 Hospira GemStar System Operating Manual ...