User Manual: HotDog
TM
Patient Warming System Controller
Page 11
INSTRUCTIONS FOR USE
The instructions below describe how to operate the Model WC5X Controller. For information
about HotDog warming devices and accessories, refer to the User Manual provided with each
device/accessory.
1. Mount the HotDog Controller on an IV pole or the OR table/gurney rail (refer to
or place the device on a flat, horizontal surface.
2. Insert the HotDog Controller power plug into a properly grounded hospital grade electrical
outlet.
WARNING: To avoid risk of electric shock, this equipment must only be connected to a
supply mains with protective earth grounding.
Note: The Controller is grounded and should not be attached to un-grounded tables
intended for use with a hyfrecator or equivalent devices.
3. After the lights illuminate in sequence, the unit will emit an audible tone and display the
software revision on the Temp Setting Display for approximately 2 seconds.
4. Position and secure the HotDog warming device (e.g., Warming Blanket, Warming
Mattress)
following instructions in the User Manual provided with the device
.
5. Insert the warming device connecting cable into the proper port on the Controller.
Controller Port
Warming Device
A and B
Warming Blanket
C
Warming Mattress
D and E
Future approved HotDog accessories
Note: When the connecting cable is inserted into the Controller, an audible beep indicates
that the control sensor and over temperature thermistor are present and functioning
properly.
6. Press the Temp Adjust Button that corresponds with the port being used until the desired
temperature is set, as indicated by the Temp Setting Display. The display will flash until
the temperature has been reached.
7. Monitor the patient’s temperature regularly. Adjust the temperature setting of the HotDog
Controller as necessary to maintain the desired patient temperature.
8. When patient warming therapy is complete, turn the Mains Power Switch to OFF.
9. After use, disconnect the HotDog Controller from the electrical outlet.
10. Discard disposable accessories following standard hospital procedure. Clean the reusable
warming device as instructed in the User Manual provided with the device.