Mounting the HotDog Controller to an IV Pole
To mount the HotDog Controller to an IV pole, place the Controller IV pole adapter around the
IV pole and turn the clamp handle clockwise until securely tightened (
). To remove the
Controller from the IV pole, turn the clamp handle counterclockwise until the unit releases.
Caution
To prevent the IV pole from tipping, the Controller must be attached at a height that provides
stability. It is recommended to use an IV pole with a minimum wheelbase radius of 35.6 cm (14
in) and to mount the Controller no higher than 112 cm (44 in) from the floor. Failure to
properly mount the Controller may result in IV pole tipping, catheter site trauma and patient
injury.
Figure 1: HotDog Controller Mounted on an IV Pole
The Controller may also be suspended from the OR table/gurney rail using optional hooks
(
).