ELECTROMAGNETIC COMPATIBILITY (EMC)
The HotDog Patient Warming System requires special precautions regarding EMC and must be
installed and put into service according to the EMC information provided in this User Manual.
Warning
Use of accessories and cables other than those specified may result in increased
emissions or decreased immunity of the HotDog Patient Warming System.
The HotDog Patient Warming System should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, carefully observe the
HotDog Patient Warming System to verify that it operates normally in this
configuration.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The HotDog™ Patient Warming System is intended for use in the electromagnetic environment specified below. The
customer or user of the HotDog Patient Warming System should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment
–
Guidance
RF emissions, CISPR 11
Group 1
The HotDog Patient Warming System uses RF energy only
for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions, CISPR 11
Class A
Harmonic emissions, IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions, IEC 61000-3-3
Complies
The HotDog Patient Warming System is suitable for use in
all establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.