User Manual: Hot Dog Veterinary Patient Warming System Controller
Page 17
Electrical Characteristics
Leakage Current
Meets UL 2601-1 and IEC 60601-1 requirements for Class I, Type BF equipment.
Power Consumption
850W maximum
Power Cord
4.6 m (15 ft)
Device Ratings
Input: 100-240 VAC, 50/60 Hz, 850VA
Output A & B: 48 VDC, 480 VA Max each
Output C: 240 VA Max
Output D & E: 48 VDC, 144 VA Max each
Fuses
T10AL250V (2 x 5x20mm)
Environmental Conditions
Environmental Conditions for Transport
and Storage
Temperature: -20°C to 60°C
Humidity: 20% to 80%
Keep Dry
Environmental Conditions for Use
Temperature: 15°C to 25°C
Humidity: 20% to 80%
Classification and Standards
Certifications
IEC 60601-1; EN 60601-1-2; UL 60601-1; CAN/CSA-C22.2, No. 601.1, EN 55011
Classification
Classified under IEC 60601-1 Guidelines (and other national versions of the
Guidelines) as Class I, Type BF, Ordinary equipment, Continuous operation. Not
suitable for use in presence of flammable anesthetic mixtures with air or with
oxygen or nitrous oxide. Classified by Intertek Testing Services NA Inc. with
respect to electric shock, fire, and mechanical hazards only, in accordance with UL
60601-1. Classified under the Medical Device Directive (93/42/EEC) as a Class IIb
device. Classified under the Canadian Medical Device Regulation as Class II.
Diagnostics
A qualified technician can perform general system testing. The Controller has no
user serviceable parts.
Important Information
This device complies with the EMC requirements according to IEC 60601-1-2.
Radio transmitting equipment, cellular phones, etc. shall not be used in the close
proximity of the device since this could influence the performances of the device.
Particular precaution must be considered during use of strong emission sources
such as High Frequency surgical equipment and similar so that, e.g., the HF-cables
are not routed on or near the device. If in doubt, contact a qualified technician or
your local representative.