User Manual: Hot Dog Veterinary Patient Warming System Controller
Page 19
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The Hot Dog™
Vet Patient Warming System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Hot Dog Patient Warming System should assure that it is used in such an
environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
–
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output
lines
±2 kV for power
supply lines
±1 kV for
input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the Hot Dog Patient Warming System
requires continued operation during power
mains interruptions, it is recommended that
the Hot Dog Patient Warming System be
powered from an uninterruptible power supply
or a battery.
Power frequency
50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.