User Manual: Hot Dog Veterinary Patient Warming System Controller
Page 16
SPECIFICATIONS
Physical Characteristics
Dimensions
33 cm high x 14.0 cm deep x 19.7 cm wide13” high x 5.5” deep x 7.75” wide
Weight
5 kg (11 lbs.)
Mounting
Can be placed on a horizontal flat surface (i.e. table top), clamped to an IV pole or
hung on a OR/gurney rail using optional hanging hooks
Temperature Characteristics
Temperature Control
Micro-processor
Operating Temperatures
Blanket Ports A and B adjustable in 1
o
C increments
37° to 43° ± 1.0°C
98.6° to 109.4° ± 1.8°F
Port C - NA
Safety System
All alarm conditions are classified as Medium Priority Technical Alarms
Primary Over-temp Alarm
Ports A and B (Warming Blanket)
Alarm sounds when temperature sensor is at set point + 1°C
Port C - NA
Secondary Over-temp Alarm
Ports A and B (Warming Blanket)
Independent electronic circuit shuts the heater off if the Warming Blanket
temperature sensor reaches set point + 3°C. (46°C)
Port C - NA
Time out timer
If warming blanket does not reach set temperature within 10 minutes the
controller will alarm
Six hour timer
If a warming blanket is left at a steady setting for six hours, the controller will
discontinue power to warming blanket.
Over-current limits
Port A
10 amps max
Port B
10 amps max
Port C
5 amps
System
14.6 amps
System Over-current Protection
Dual input fused lines.