Use of compressed air for emptying
ATTENTION! When using compressed air, the measuring components can be destroyed!
Measuring ranges of the components in the measuring box:
Measuring medium
Water
Max. medium temperature
85 °C (short-term 95 °C)
Max. operating pressure at medium temperature
over the service life 12 bar at +40 °C
over the service life 8 bar at +70 °C
Pressure measurement
Measuring range
0
–
12 bar
Resolution
0.1 % of measuring range end
Measuring accuracy
0.5 % of measuring range end
Flow rate
Measuring range
1.8
–
32 l/min
Measuring principle
Piezoceramic sensor element
Measuring accuracy at < 50 % FS* (water)
< 1 % FS*
Measuring accuracy at > 50 % FS* (water)
< 2 % measured value
* Full scale (final value)
Temperature measurement
Measuring principle
Resistance Pt1000
Measuring accuracy
± 0.3 K