D-System Operating Instructions
Hotset GmbH
–
Hueckstraße 16
–
58511 Lüdenscheid, Germany
–
+49 2351 4302 0
–
info@hotset.com
–
–
04/2021 RS
14
Connection 2 or the arrow points out of the circle
The medium comes out of the measuring module at this connection point and can be
connected from here to a measurement object, such as an injection molding tool, heat
exchanger, etc.
It is also possible to make a direct connection from port 2 to port 3. This allows, for example,
pump capacities to be determined, or temperature control units to be calibrated.
Tip: Maximum pressure and flow values of the supply (observe maximum measuring range!) can
be tested if it is possible to integrate a valve for flow regulation in the hose connection from port
2 to port 3.
Connection 3 or the arrow points into the circle
If the medium has flowed through the consumer/measurement object, it is connected to port 3.
The output temperature T2, output pressure p2 and flow rate q are measured in the measuring
module.