TM 11-6625-3071-14
3478A
4-99. AC Currant Calibration and Test (Optional)
4-100. Since the DC Current and AC Volts Calibration
has been performed and the AC Current Performance
Test passed, the AC Current Calibration and Test need
not to be done. If there is any doubt about the accuracy
of the AC Current Function, the function can be
calibrated and checked using the Optional AC Current
Calibration and Test Procedures in the following
paragraphs. Refer to Table 4-11 for the AC Current
Calibration Signals, Performance Test Signals, and
Performance Test Limits. Each step in the table is
shown in parenthesis in the procedure. All test signals
are applied
Table 4-11. AC Current Calibration and Test
4-101. Equipment Required. The following is the
required test equipment for the AC Current Calibration
and Test.
AC Calibrator (Fluke Model 5200A Option 05)
AC-DC Current Calibrator (Valhalla Model 2500)
4-102. Calibration and Test Procedure. Do the
following:
a. Reset the 3478A by pressing the blue Shift button
and then the SGL/TRIG (TEST/RESET) button.
b. Make sure the front panel CAL ENABLE Switch is
set to the calibration enable position (the slot of the
switch is in the vertical position).
c. (Step #1.) Set the 3478A to the AC Current
Function and the 3OOmA Range.
d. Set the multimeter to the Single Trigger Mode by
pressing the SGL/TRIG button.
e. Set the AC Calibrator for a IV at 1KHz output.
f. Set the AC-DC Current Calibrator to the 100mA
Range.
g. Connect the AC-DC Current Calibrator to the AC
Calibrator as shown in Figure 4-5.
h. Connect the 3478A to the AC-DC Calibrator as
shown in Figure 4-5.
i. Press the blue Shift button and then the LOCAL
(CAL) button.
j. Using the 3478A’s Uprange and/or Downrange
buttons, adjust the reading on the multimeter until it
agrees with the output value of the AC-DC Current
Calibrator (if the output value is known).
k. Press the SGL/TRIG button once.
l. This completes the AC Current Calibration. If a
Performance Test is desired, continue with the next step.
Otherwise remove the 3478A and the AC Calibrator from
the AC-DC Current Calibrator.
m. (Step #2.) If a Performance Test is desired, set the
AC Calibrator for a IV at 5KHz output.
n. Set the AC-DC Current Calibrator to the O1mA
Range. Check the reading on the 3478A.
o. (Step #3.) Set the AC-DC Current Calibrator to the
100mA Range. Check the reading.
p. (Step #4.) Set the 3478A to the 3A Range.
q. Set the AC-DC Current Calibrator to the IA Range.
Check the reading.
r. Remove the 3478A and the AC Calibrator from the
AC-DC Current Calibrator. This completes the AC
Current Calibration and Test.
4-103. Ohms Calibration and Test
4-104. The 3478A’s Ohms Function can be calibrated in
either the 2-Wire or 4-Wire Ohms Function. It cannot be
calibrated in both. Make sure the multimeter is
calibrated in the function in which it will be used. If the
3478A is to be calibrated in the 2-Wire Ohms Function,
connect the Standard Resistors to the 3478A’s HI and
LO INPUT Terminals. If the 3478A is to be calibrated in
the 4-Wire Ohms Function, connect the Standard
Resistors as shown in Figure 4-7.
4-105. Refer to Table 4-12 for the Ohms Calibration
Signals, Performance Test Signals, and Performance
Test Limits. Each step in the table is shown in
parenthesis in the procedure. All test signals are applied
to the INPUT Terminals.
4-106. Equipment Required. The following is the
required test equipment for the Ohms Calibration and
Test.
10 ohm
±
.005%; (Guildline Model 9330/10 or
9330A/10)
100 ohm
±
.005%; (Guildline Model 9330/100 or
9330A/100
1K ohm
±
.005%; (Guildline Model 9330/IK or
9330A/IK)
10K ohm
±
.001%; (Guildline Model 9330/10K or
9330A/10K)
100K ohm ± .001%; (Guildline Model 330/100K or
4-18
Summary of Contents for 3478A
Page 2: ...TM 11 6625 3071 14 A ...
Page 4: ...TM 11 6625 3071 14 C D BLANK ...
Page 12: ...TM 11 6625 3071 14 Table 1 1 Specification 1 2 ...
Page 13: ...TM 11 6625 3071 14 Table 1 1 Specifications Cont 1 3 ...
Page 14: ...TM 11 6625 3071 14 Table 1 1 Specifications Cont 1 4 ...
Page 53: ...TM 11 6625 3071 14 1 ...
Page 54: ...TM 11 6625 3071 14 2 ...
Page 55: ...TM 11 6625 3071 14 3 ...
Page 56: ...TM 11 6625 3071 14 4 ...
Page 87: ...TM 11 6625 3071 14 3478A Figure 7 D 3 Flowchart B 7 D 5 ...
Page 88: ...TM 11 6625 3071 14 3478A Figure 7 D 4 Flowchart C 7 D 6 ...
Page 91: ...TM 11 6625 3071 14 3478A Figure 7 D 6 Flowchart D 7 D 9 ...
Page 111: ...TM 11 6625 3071 14 Figure 7 F 17 3478A Simplified Reference Circuitry 7 F 16 ...
Page 122: ...TM 11 6625 3071 14 Table A 2 A 2 HP IB Worksheet A 4 ...
Page 137: ...TM 11 6625 3071 14 Figure 7 D 2 Flow chart A 7 D 3 ...
Page 139: ...TM 11 6625 3071 14 Figure 7 G 2 3478A Block Diagram 7 G 3 ...
Page 140: ...TM 11 6625 3071 14 Component Locator for Input Circuitry and Ohms Current Source 7 G 4 ...
Page 141: ...TM 11 6625 3071 14 Figure 7 G 3 Input Circuitry and Ohms Current Source 7 G 5 ...
Page 142: ...TM 11 6625 3071 14 F G 6 ...
Page 143: ...TM 11 6625 3071 14 2 Figure 7 G 4 AC to DC Converter 7 G 7 ...
Page 144: ...TM 11 6625 3071 14 Component Locator for A D Converter and Control Logic 7 G 8 ...
Page 145: ...TM 11 6625 3071 14 3 Figure 7 G 5 A D Converter and Control Logic 7 G 9 ...
Page 146: ...TM 11 6625 3071 14 7 G 10 ...
Page 147: ...TM 11 6625 3071 14 4 Figure 7 G 6 Power Supplies 7 G 11 7 G 12 blank ...
Page 148: ......
Page 149: ...PIN NO 057444 ...