Chapter 4 Command Reference
Introduction
Programming Guide
4-33
:DIAGnostic:CALibration:INPut[1|2]:OFFSet:
AUTO ONCE | OFF
Calibrates the channel 1 or 2 input trigger OFFSet when the ONCE parameter is
used.
Before sending this command, BE SURE to disconnect any input signal from the
appropriate input.
A sequence of ASCII-encoded bytes: OFF
•
The calibration values are stored in non-volatile memory, and are unaffected
by power-on, save/recall, and *RST.
•
Use :DIAG:CAL:STAT? to check for successful calibration.
•
After calibration is completed, the state of this command’s parameter is OFF.
Scale and Offset/POWER (Calibration menu).
:DIAGnostic:CALibration:INTerpolator:AUTO ONCE | OFF |ON
Calibrates the interpolator circuit in the Counter when the ONCE parameter is
used.
AUTO ON enables automatic interpolator calibration on every measurement.
AUTO OFF disables automatic interpolator calibration.
A sequence of ASCII-encoded bytes: OFF or ON
•
*RST: ON
•
Use :DIAG:CAL:STAT? to check for successful calibration.
•
After ONCE calibration is completed, the state of this command’s parameter
is OFF.
•
When :DIAG:CAL:INT is set to OFF, the Counter reports the positive status
condition (questionable Time, Frequency, and Phase) to bits 2, 5, and 6 of the
Questionable Status Register. When :DIAG:CAL:INT is set to ON, the
Counter reports the negative status condition (NOT questionable Time,
Frequency, and Phase) to bits 2, 5, and 6 of the Questionable Status Register.
•
This enable is unaffected by save/recall.
Query Response
Comments
Related
Front-Panel
Keys
Query Response
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...