Chapter 4 Command Reference
:DISPlay Subsystem
4-38
Programming Guide
:DISPlay[:WINDow]:TEXT:FEED “[:]CALCulate2” |
“[:]CALCulate3”
Sets or queries what data flow is fed into the display.
Choose from the following <data_handle> strings:
•
“[:]CALCulate2”— should be used to direct any result
other than the
statistics
to the result display
Specifically this would select one of the following results for the result
display:
– raw measurement (if math is disabled— :CALC:MATH:STAT OFF)
– the scaled/offset measurement (if math is enabled— :CALC:MATH:STAT
ON)
– the limit graph (if limit testing is enabled with the graphic display—
:CALC2:LIM:STAT ON, :CALC2:LIM:DISP GRAPh)
•
“[:]CALCulate3”— should be used to direct the statistical result
(if statistics are enabled, :CALC3:AVER[:STAT] ON) to the result display;
the particular statistic displayed is determined by :CALC3:AVER:TYPE.
A string is returned: “CALC2” or “CALC3.”
•
*RST: “:CALCulate2”
•
Refer to the section titled “How to Program the Counter to Display Results”
in Chapter 3 of this guide.
Stats
Query Response
Comments
Related
Front-Panel
Keys
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...