Chapter 4 Command Reference
Introduction
Programming Guide
4-85
[:SENSe]:EVENt3:SLOPe?
Queries which edge of channel 3 input port will be considered an event.
A sequence of ASCII-encoded bytes: POS
[:SENSe]:FREQuency Subtree
This subtree controls the Frequency, Frequency Ratio, and Period measuring
capabilities of the instrument.
[:SENSe]:FREQuency:ARM Subtree
This subtree is used to synchronize the Frequency, Frequency Ratio, and Period
start and stop arm with events. The following combination of start/stop arming
sources are valid:
STARt:SOURce
STOP:SOURce
Front-Panel Gating Settings
GATE
STOP
IMMediate
IMMediate
AUTO
_______
IMMediate
TIMer
TIME
_______
IMMediate
DIGits
DIGITS
_______
EXTernal
EXTernal
EXTERNAL
POS, NEG
EXTernal
TIMer
EXTERNAL
TIME
EXTernal
IMMediate
EXTERNAL
AUTO
[:SENSe]:FREQuency:ARM[:STARt]:SLOPe POSitive | NEGative
Sets or queries the slope of the external start arm signal used in external arming
Frequency, Frequency Ratio, and Period measurements.
A sequence of ASCII-encoded bytes: POS or NEG
•
*RST: POSitive
•
Only applies when [:SENS]:FREQ:ARM[:STAR]:SOUR EXT is selected.
Gate & ExtArm
Query Response
Query Response
Comments
Related
Front-Panel
Keys
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...