Chapter 4 Command Reference
[:SENSe] Subsystem
4-106
Programming Guide
[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate |
EXTernal TIMer
Sets or queries the stop arm for Totalize measurements.
A sequence of ASCII-encoded bytes: IMM, EXT or TIM
•
*RST: TIMer
•
When both start and stop arm are set to IMMediate, use :INIT[:IMM] to
begin totalizing and :ABORt to terminate. No valid totalize result will exist
until the “auto” totalize measurement is terminated.
Gate & ExtArm
[:SENSe]:TOTalize:ARM:STOP:TIMer
<numeric_value> [S]
Sets or queries the gate time used in arming Totalize measurements.
•
For short gate time: 1E-3 to 99.99E-3 seconds
•
For long gate time: 10 0E-3 to 1000.000 seconds
•
For short gate time: 0.01E-3 seconds
•
For long gate time: 1E-3 seconds
Numeric data transferred as ASCII bytes in <NR3> format with six significant
digits.
•
*RST: 100E-3 S
•
Only applies when [:SENS]:TOT:ARM:STOP:SOUR TIM is selected.
Gate & ExtArm
Query Response
Comments
Related
Front-Panel
Keys
<numeric_value>
Range
<numeric_value
Resolution
Query Response
Comments
Related
Front-Panel
Keys
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...