Chapter 4 Command Reference
*SAV <NRf>
(Save Command)
4-138
Programming Guide
*SAV <NRf>
(Save Command)
*SAV <NRf>
This command stores the current state of the instrument in local
non-volatile memory .
The current instrument state is saved in register 0 when *RCL or
front-panel recall is executed.
1 to 20
1
The following states are
not
saved:
*EMC
*ESE
*OPC
*SRE
:CALibration:COUNt?
:CALibration:DATA
:CALibration:SECurity:CODE
:CALibration:SECurity:STATe
:CONFigure? response
:DIAGnostic:CALibration:INTerpolator:AUTO
:DISPlay:ENABle
:DISPlay:MENU[:STATe]
:DISPLay[:WINDow]:TEXT:RADix
:FETCh? implied function
:READ? implied function
[:SENSe]:FREQuency:EXPected[1|2|3]
[:SENSe]:FREQuency:EXPected[1|2|3]:AUTO
[:SENSe]:ROSCillator:EXTernal:CHECk
[:SENSe]:ROSCillator:SOURCe
[:SENSe]:ROSCillator:SOURCe:AUTO
:STATus:OPERation:ENABle
:STATus:OPERation:NTRansition
:STATus:OPERation:PTRansition
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:NTRansition
:STATus:QUEStionable:PTRansition
:SYSTem:COMMunicate:SERial:CONTrol:DTR
:SYSTem:COMMunicate:SERial:TRANsmit:BAUD
:SYSTem:COMMunicate:SERial:TRANsmit:PACE
:SYSTem:COMMunicate:SERial:TRANsmit:PARity[:TYPE]
HP-IB Address
Error Queue
Key Queue
<NRf> Range
<NRf>
Resolution
Comments
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...