Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Programming Guide
2-29
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax
Parameter Form
Std/
New
Comments
:INITiate
:AUTO
:CONTinuous
[:IMMediate]
<Boolean>
<Boolean>
Std
New
Std
Std
Subsystem. Controls the initiation of measurements.
AUTO ON enables the Counter to automatically stop
measuring on a limit test failure. AUTO OFF
disables the automatic stop.
Sets the enable for continuously initiated
measurements.
Event; no query. Causes the instrument to initiate
the number of measurements specified by
:TRIGger:COUNt:AUTO.
:INPut[1|2]
:ATTenuation
:COUPling
:FILTer
[:LPASs]
[:STATe]
:FREQuency?
:IMPedance
1 | 10
AC | DC
<Boolean>
<numeric_value>
[OHM]
Std
Std
Std
Std
Std
Std
Std
Std
Subsystem. Controls the characteristics
of the instrument ’s input ports. :INPut1= channel 1
and :INPut2= channel 2
Sets input attenuation.
Sets input coupling.
Subtree. Allows a low pass filter to be inserted in the
path of the measurement signal.
Subtree. Controls the Low PASs filter.
Sets the Low PASs filter enable.
Query only. Returns the cutoff frequency of the low
pass filter. Units are Hertz.
Sets input impedance (50
Ω
or 1 M
Ω
).
:INPut3
:COUPling?
:IMPedance?
Std
Std
Std
Subsystem. Queries the characteristics of the
Counter’s input channel 3.
Query only. Returns channel 3 input coupling.
Query only. Returns channel 3 input impedance.
:MEASure
Std
See Measurement Instructions in this table.
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...