Chapter 2 Command Summary
HP 53131A/132A Command Summary
2-36
Programming Guide
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax
Parameter Form
Std/
New
Comments
[:SENSe] (Cont.)
:TOTalize (Cont.)
:ARM (Cont.)
:STOP
:SLOPe
:SOURce
:TIMer
POSitive | NEGative
EXTernal | TIMer | IMMediate
<numeric_value> [S]
New
New
New
New
Subtree.
Sets the slope of the external stop arm signal used
in external arming totalize measurements. Only
applies when [:SENS]:TOT:ARM:STOP:SOUR
EXT is selected.
Sets the stop arm for totalize measurements.
Sets the gate time used in totalize measurements.
Only applies when
[:SENS]:TOT:ARM:STOP:SOUR TIM is selected.
:STATus
:OPERation
:CONDition?
:ENABle
[:EVENt]?
:NTRansition
:PTRansition
:PRESet
:QUEStionable
:CONDition?
:ENABle
[:EVENt]?
:NTRansition
:PTRansition
<non-decimal numeric> |
<NRf>
<non-decimal numeric> |
<Nrf>
<non-decimal numeric> |
<NRf>
<non-decimal numeric> |
<NRf>
<non-decimal numeric> |
<NRf>
<non-decimal numeric> |
<NRf>
Std
Std
Std
Std
Std
Std
Std
Std
Std
Std
Std
Std
Std
Std
Subsystem. Controls the SCPI-defined (Operation
and Questionable) status-reporting structures.
Subtree.
Query only. Queries the Operation Condition
Status Register.
Sets the Operation Event Status Enable Register.
Query only. Queries the Operation Event Status
Register.
Sets the negative transition filter for the Operation
status reporting structure.
Sets the positive transition filter for the Operation
status reporting structure.
Event; No query. Presets the enable registers and
transition filters associated with the Operation and
Questionable status reporting structures.
Subtree.
Query only. Queries the Questionable Data
Condition Status Register.
Sets the Questionable Data Event Status Enable
Register.
Query only. Queries the Questionable Data Event
Status Register.
Sets the positive transition filter for the
Questionable Data status reporting structure.
Sets the negative transition filter for the
Questionable Data status reporting structure.
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...