Chapter 2 Command Summary
*RST Response
2-40
Programming Guide
*RST Response
The IEEE 488.2 *RST command returns the instrument to a specified state
optimized for remote operation. (Use *CLS to clear the status event registers and
the SCPI error queue.)
The states of commands affected by the *RST command are described in Table 2-
3. Since the HP 53131A and HP 53132A have different arm ing capabilities for
Time Interval measurements , sub-tables (Table 2-3A and Table 2-3B) are
provided. Table 2-3A describes the state of the HP 53131A’s [:SENSe]:Interval
commands affected by the *RST command , and Table 2-3B describes the
HP 531312A’s [:SENSe]:Interval commands affected by the *RST command .
Table 2-4 lists commands that are unaffected by *RST.
Table 2-3. HP 53131A /132A *RST State
Command Header
Parameter
State
:CALCulate[1]:FEED
:CALCulate[1]:IMMediate:AUTO
:CALCulate[1]:MATH[:EXPRession]:NAME
:CALCulate[1]:MATH:STATe
“[:]SENSe[1]”
<Boolean>
SCALE_OFFSET
<Boolean>
“SENSe[1]”
OFF
SCALE_OFFSET
OFF
:CALCulate2:FEED
:CALCulate2:IMMediate:AUTO
:CALCulate2:LIMit:CLEar:AUTO
:CALCulate2:LIMit:DISPlay
:CALCulate2:LIMit:LOWer[:DATA]
:CALCulate2:LIMit:STATe
:CALCulate2:LIMit:UPPer[:DATA]
“[:]CALCulate[1] ”
<Boolean>
<Boolean>
GRAPh | NUMBer
<numeric_value>
<Boolean>
<numeric_value>
“CALCulate[1] ”
OFF
ON
NUMBer
0.0000000000
OFF
0.0000000000
:CALCulate3:AVERage:COUNt
:CALCulate3:AVERage[:STATe]
:CALCulate3:AVERage:TYPE
:CALCulate3:FEED
:CALCulate3:LFILter:LOWer[:DATA]
:CALCulate3:LFILter:STATe
:CALCulate3:LFILter:UPPer[:DATA]
<numeric_value>
<Boolean>
MAXimum | MINimum |
SDEViation | MEAN
“[:]CALCulate[1] ”
<numeric_value>
<Boolean>
<numeric_value>
100
OFF
MEAN
“CALCulate[1] ”
0.0000000000
OFF
0.0000000000
*
DDT
<arbitrary block>
#14INIT
:DIAGnostic:CALibration:INTerpolator:AUTO
ON | OFF | ONCE
ON
:DISPlay:ENABle
:DISPlay:MENU[:STATe]
:DISPlay[:WINDow]:TEXT:FEED
<Boolean>
OFF
“[:]CALCulate2 ” |
“[:]CALCulate3 ”
ON
OFF
“CALCulate2”
Summary of Contents for 53131A
Page 1: ...Programming Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 13: ...1 1 Before You Start ...
Page 24: ...Chapter 1 Before You Start Related Documentation 1 12 Programming Guide ...
Page 25: ...2 2 Command Summary A Quick Reference ...
Page 68: ...Chapter 2 Command Summary RST Response 2 44 Programming Guide ...
Page 69: ...3 3 Programming Your Universal Counter for Remote Operation ...
Page 167: ...4 4 Command Reference A Dictionary ...
Page 310: ...Chapter 4 Command Reference WAI Wait to Continue Command 4 144 Programming Guide ...
Page 311: ...5 5 Errors ...
Page 329: ......
Page 330: ...H Manual Part Number 53131 90044 Printed in U S A NOVEMBER 1996 ...