Chapter 2 Service
The HP 53131A/132A Calibration Procedures
Assembly-Level Service Guide
2-19
2
8
To calibrate the gain for Channel 2, connect the input or the
BNC Tee to Channel 2 of the Counter.
9
Press any one of the arrow keys until
CAL: GAIN 2 ?
is displayed.
10
Press Enter key.
A scrolling message is displayed. (The +5V dc supply should already be
connected to Channel 2 as you were previously instructed to do this in
step 8.)
11
Press Enter key again.
The Counter momentarily displays
CALIBRATING
, and then it should
display
GAIN 2 PASS
.
If the fail message is displayed, refer to the troubleshooting section in this
chapter.
12
Disconnect the calibration setup.
This completes the Gain Calibration procedure. Proceed to the next
procedure to calibrate the time interval.
To Calibrate Time Interval
Refer to the section titled “INFORMATION ABOUT CAL:TI QUIK?
AND CAL:TI FINE?” (page 2-10 and page 2-11) for details on the
differences between the two time interval calibrations.
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...