Chapter 2 Service
The HP 53131A/132A Calibration Procedures
Assembly-Level Service Guide
2-21
2
5
Press Enter key.
The Counter displays
CALIBRATING
, and then it should display
TI CAL PASS
.
If the fail message is displayed, refer to the troubleshooting section in this
chapter.
6
Disconnect the calibration setup.
This completes the TI QUIK? calibration procedures.
Depending on which timebase the Counter contains, proceed to the
procedure “To Calibrate the Standard Timebase” (page 2-23) or the
“To Calibrate the High Stability Timebase Option (Medium, High, or
Ultra-High)” procedure (page 2-24) to calibrate the timebase.
CAL: TI FINE? Calibration
1
Connect the HP 8130A Pulse Generator, HP 59992A J06 Time
Interval Calibrator, and HP 53131A/123A Counter as shown in
Figure 2-4.
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 59992A J06 Time Interval Calibrator (or equivalent)
HP 10503A BNC Cable (3)
Figure 2-4. Calibration Setup for CAL: TI FINE?
HP 53131A/132A
Counter
1
2
3
4
In
A
B
HP 59992A J06
Time Interval Calibrator
Start
Stop
HP 8130A
Pulse Generator
Output
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...