Chapter 2 Service
About the HP 53181A Calibration Menu
2-26
Assembly-Level Service Guide
2
•
unsecure for calibration by entering the security code while the
Counter is secured (CODE: )
•
change the security code by entering a new code while the Counter is
unsecured (CODE: )
•
view the calibration count, which indicates the number of times that
the Counter has been calibrated (CAL COUNT?). The calibration
count increments up to a maximum of 32,767 after which it wraps
around to 1. A calibration count of 0 indicates that the Counter is
completely uncalibrated. (The calibration count is unaffected by
interpolator calibration.)
•
get Help (HELP:) with any of the following:
–
how to calibrate (CAL?),
–
how to secure (SECURE?), and
–
how to change the code (CODE?)
NOTE
A good precaution is to use an HP-IB program (see the programming
example titled “To Read and Store Calibration Data” in Chapter 3 of the
Programming Guide) to read and store the calibration factors prior to
initiating any calibration.
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...