Chapter 1 Performance Tests
Test 1: Time Interval (HP 53131A/132A Only)
1-18
Assembly-Level Service Guide
1
Procedure
1
Connect the pulse generator signal to Channel 1 of the Counter as
shown in Figure 1-5.
Figure 1-5. Time Interval (Channel 1 TO 2) Test Setup
2
Set the Pulse Generator to the following:
PERIOD: 10
µ
s
WIDTH: 150 ns
High: 1.00 V
Low : 0.00 V
Input Mode: Normal
3
On the Counter, Press Single/Stop key.
The HP 53131A Counter should display 0.1500
µ
s ±0.0040
µ
s.
The HP 53132A Counter should display 0.1500
µ
s ±0.0025
µ
s.
4
Record this value in the appropriate place in the Performance
Test Record (Test 1, Line 1 for HP 53131A or Test 1, Line 2 for
HP 53132A).
5
Disconnect the test setup.
HP53131A/132
Counter
Channel 1
HP 8130A
Pulse Generator
Output
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...