Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
Assembly-Level Service Guide
1-31
1
100 MHz to 2.5 GHz Sensitivity for HP 53131A/132A
Option 030 only
1
Set the HP 8663B Synthesized Signal Generator to output a
2500 MHz,
−
40 dBm sine wave, and connect the signal to Channel 3
of the Counter as shown in Figure 1-9.
Figure 1-9. Option 030/050/124 Channel 3 Frequency Sensitivity and Accuracy Test Setup
2
On the Counter, Press Run key.
HP 53131A/132A
Counter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Channel 3
50
Ω
Feedthrough
Ref. In
(rear panel)
Input
(rear panel)
HP 8340B
Sweep Oscillator
Frequency Standard
(switch set to EXT)
Output
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...