Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
1-48
Assembly-Level Service Guide
1
Test 1: Trigger Level (HP 53181A Only)
This test verifies the Trigger Level accuracy of the HP 53181A Frequency
Counter.
Equipment
HP 3325B Synthesizer
HP 10503A BNC Cable
Counter Setup
1
Cycle the POWER key to preset the Counter.
2
Set CHANNEL 1 input condition to:
Trigger/Sensitivity,
AUTO TRG: OFF
Trigger/Sensitivity,
LEVEL: 0.000V
50
Ω
/1M
Ω
, 50
Ω
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...