1
Connect the equipment as shown in Figure 1-13 with the signal
generator connected to Channel 1 of the Counter.
Figure 1-13. Channels 1 Frequency Accuracy Test Setup
2
Set the HP 8663A Signal Generator to output 1 MHz at
−
7 dBm.
3
On the Counter, press Run key.
Verify the Counter reads 1 MHz ±1 mHz.
HP 53181A
Counter
Channel 1
50
Ω
Feedthrough
Ref. In
(rear panel)
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Input
(rear panel)
HP 8340B
Sweep Oscillator
Frequency Standard
(switch set to EXT)
Output