Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
Assembly-Level Service Guide
1-59
1
100 MHz to 1.5 GHz Sensitivity for Option 015 Only
1
Set the HP 8663B Synthesized Signal Generator to output a
500 MHz,
−
60 dBm sine wave and connect the signal to Channel 2
of the Counter as shown in Figure 1-14.
Figure 1-14. Option 015/030/050 Channel 2 Frequency Test
2
On the Counter, Press Run key.
HP 53181A
Counter
Channel 2
50
Ω
Feedthrough
Ref. In
(rear panel)
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Input
(rear panel)
HP 8340B
Sweep Oscillator
Frequency Standard
(switch set to EXT)
Output
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...