Chapter 2 Service
About the HP 53131A/132A Calibration Menu
2-8
Assembly-Level Service Guide
2
The HP 53131A/132A Calibration Menu Tree
NOTE
Turn power off, press and hold Scale & Offset key, then press POWER
key to access this menu. This menu does not exist in early versions
(firmware revisions 3317, 3335, and 3402) of the HP 53131A/132A.
In the early versions of the Counter, the CAL: menu item resides
in the Utility menu, and there is no calibration security
capability. See Chapter 6, “Backdating,” in this guide for correct
calibration procedures.
1
This menu item appears and calibration is permitted only if calibration is unsecure. Enter in the correct code
to change calibration to secure.
2
Timebase can be automatically calibrated only if the timebase option is installed.
Scale &
Offset
On / Stby
POWER
CAL SECURE
CAL UNSECURE
1
CAL:
OFFS2?
CAL: OFFS1?
CAL:
GAIN2?
CAL:
GAIN1?
CAL:
TIMEBAS?
CAL:
TI FINE?
CAL:
TI QUIK?
CODE: 0
CAL COUNT?
HELP: CAL?
HELP:
SECURE?
HELP:
CODE?
2
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...